Abstract

Purpose

To examine the effect of cannula diameter and conjunctival flap method on bleb survival in rabbits undergoing cannula-based glaucoma filtration surgery (GFS).

Methods

Twelve New Zealand White rabbits underwent GFS in both eyes. The twenty-four eyes were divided into four groups. Two of the four groups (N = 12) received limbus-based con- junctival flaps (LBCF), and the other two (N = 12) received fornix-based conjunctival flaps (FBCF). Six FBCF rabbit eyes were implanted with 22-gauge drainage tubes, and the other six were implanted with 26-gauge tubes. Likewise, six LBCF rabbits received 22-gauge drainage tubes and six received 26-gauge tubes. Filtration blebs were evaluated every three days by a masked observer. Bleb failure was defined as the primary endpoint in this study and was recorded after two consecutive flat bleb evaluations.

Results

Group 1 (LBCF, 22- gauge cannula) had a mean bleb survival time (Mean ± SD) of 18.7 ± 2.9 days. Group 2 (LBCF, 26-gauge cannula) also had a mean bleb survival time of 18.7 ± 2.9 days. Group 3 (FBCF, 22-gauge cannula) had a mean bleb survival time of 19.2 ± 3.8 days. Group 4 (FBCF, 26-gauge cannula) had a mean bleb survival time of 19.7 ± 4.1 days. A 2-way analysis of variance showed that neither surgical approach nor cannula gauge made a statistically significant difference in bleb survival time (P = 0.634 and P = 0.874). Additionally, there was no significant interaction between cannula gauge and conjunctival flap approach (P = 0.874), suggesting that there was not a combination of drainage gauge and conjunctival flap method that produced superior bleb survival.

Conclusion

Limbus and fornix-based conjunctival flaps are equally effective in promoting bleb survival using both 22 and 26-gauge cannulas in the rabbit model. The 26-gauge drainage tube may be preferred because its smaller size facilitates the implantation process, reducing the risk of corneal contact.

Source Citation

Rodgers CD, Meyer AM, Rosenberg NC, Lukowski ZL, Schaefer JL, Martorana GM, et al. (2018) The impact of conjunctival flap method and drainage cannula diameter on bleb survival in the rabbit model. PLoS ONE 13(5): e0196968. https:// doi.org/10.1371/journal.pone.0196968

Comments

© 2018 Rodgers et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.